4. Alarm
Alarm Indication
The alarm condition is indicated by visual ALARM
INDICATOR as described in Table 3 below. The operator
could be patient, caretaker or medical personnel.
Table 3. Characteristics of ALARM INDICATOR
Alarm
Priority
Color
Flashing
Frequency
Medium
YELLOW
0.4Hz
Low
BLUE
Constant (ON)
Alarm Limits
Two types of alarm are present, physiological and technical
(refer to Table 4 and Table 5 below for details). The flashing
ALARM INDICATOR occurs when the alarm condition is
exceeded.
Table 4. Physiological Alarm Medium Priority
Parameter
Ranges
Defaults
SpO
2
High
Limit
86~100%, with 1% step size
Off