85~95%, with 1% step size
75~250bpm, with 5bpm step size
40~110bpm, with 5bpm step size
Note:
1. The low limit must be set below the high limit setting.
While the high limit is set equal to or below the low
limit, the low limit will automatically be adjusted to the
next setting below the newly entered high limit setting.
2. When default alarm setting is changed, a decimal point
(.) is displayed after the reading as a reminder.
3. When PR/SpO
2
high (or low) limit is exceeded,
Medium priority alarm is activated.
The PR/SpO
2
number on the display turns
YELLOW.
Table 5. Technical Alarm – Low Priority
Remaining battery is less than 25%
Poor signal, determined by the
algorithm, for more than 25 seconds
NOTE:
1 Alarms are prioritized so that high priority alarms