82
Table 7. Recommended Separation Distances
This table details the recommended separation distances
between portable and mobile RF communications equipment
and the oCare
TM
Pro 100.
This device is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled.
Users of this device can help prevent electromagnetic
interference by maintaining a minimum distance between
portable and mobile RF communication equipment
(transmitters) and the device as recommended below,
according to maximum output power of the communications
equipment.
Separation Distance According to
Frequency of Transmitter
Rated Maximum
Output Power of
Transmitter, W
150 kHz to
80 MHz
d = 1.17
P
80 MHz to
800 MHz
d = 1.17
P
80 MHz to
800 MHz
d = 2.33
P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.73
1
1.2
1.2
2.3
10
3.7
3.7
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed
above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency