76
Type of Protection
Degree of Protection
Mode of operation
Internally Powered (Battery Power)
Type BF-Applied Part
Continuous
Alarm
Alarm Type
Low Battery
Malfunction
Poor Signal Condition
High/Low SpO
2
(PR)
High/Low SpO
2
(PR) alarm range
SpO
2
high limit: 86~100 %SpO
2
SpO
2
low limit: 85~95 %SpO
2
PR high limit: 75~250 bpm
PR low limit: 40~110 bpm
Medium priority
ALARM INDICATOR: YELLOW
Alarm flashing frequency: 0.4 Hz
Low priority
ALARM INDICATOR: BLUE
Alarm flashing frequency: Constant (ON)