Table 9. Electromagnetic Immunity
This oCare
TM
Pro 100 is intended for use in the
electromagnetic environment specified below. The customer
and/or user of the oCare
TM
Pro 100 should ensure that it is used
in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
±8 kV
contact
±15 kV air
±8 kV
contact
±15 kV air
Floors should be
wood, concrete,
or ceramic tile. If
floors are covered
with synthetic
material, relative
humidity should
be at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for
power
supply lines
±1 kV for
input/output
lines
N/A
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
Surge
IEC 61000-4-5
±1 kV
differential
mode
±2 kV
common
N/A
Mains power
quality should be
that of a typical
commercial or
hospital