Performance Characteristics
Table
6-1
2465A/2455A/2445A Electrical Characteristics
Characteristics Performance Requirements
VERTICAL
DEFLECTION SYSTEM-CHANNEL 1
AND
CHANNEL 2
Deflection Factor
Range 2 mV/division to
S V/division
in
a 1-2-S
sequence
of
11
steps.
Accuracy 1
Mn
input, noninverted.
+1SoC
to
+3SoC
On-Graticule Accuracy Within ± 2%
at
any VOL TS/DIV setting for a
four
or
five-division signal centered on the
screen.
Do
V Accuracy (using ±(1.2S% of reading
+0.03
div + signal
cursors over entire aberrations).
graticule area)
-1SoC
to
+1SoC
and
Add ± 2%
of
reading.
+3SoC
to
+SsoC
SO
n Coupling Add ± 1 %
of
reading.
CH
2 Inverted Add ± 1 %
of
reading.
t:N Range
± 8 X
VOL TS/DIV setting.
V/DIV VARiable, non inverted
Continuously variable between
VOL TS/DIV
settings. Extends deflection factor
to
> 12.S
V/division.
Frequency Response Bandwidth is measured with a leveled, low
distortion,
so-n source, sine-wave generator,
terminated
in
SO
n.
The reference signal
amplitude is set at the lesser of 6 divisions
or
the maximum leveled amplitude.
Bandwidth with probe is checked using a
probe-tip-to-GR
(017
-OS20-00)
termination
adapter.
Bandwidth with external termination
is
checked
using a BNC
so-n feed through terminator
(011-0049-01
).
6-2
REV FEB 1987
246SA/24SSA/244SA Operators