EasyManuals Logo

Tektronix 455 User Manual

Tektronix 455
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
Adjustments—455/A2/B2
C. HORIZONTAL MODULE
Equipment Required
1. Time-Mark Generator
5. 50 £2 Termination
2. Sine-Wave Generator 6. Low-Capacitance Screwdriver
3. Amplitude Calibrator 7. Screwdriver
4. 50 £2 Cable
Preset front panel controls as follows:
NOTE
Do not preset internal controls.
Crt
INTENSITY Midrange
Horizontal
HORIZ DISPLAY A INTEN
TRIG MODE
AUTO
X10MAG out (off)
A COUPLING
AC
A SOURCE
NORM
ATIM E/DIV
1 ms
B TIME/DIV 5 ps
VARTIM E/DIV Calibrated (detent)
position
B SOURCE STARTS AFTER
DELAY
Set all other controls as desired.
The oscilloscope should produce a baseline trace with an
intensified portion when the controls are set as above. Ad
just INTENSITY and FOCUS controls as needed to main
tain a well-defined display while making adjustments.
Cl. SWEEP START-STOP
a. Connect 1 ms markers from time-mark generator to
vertical input via 50 £2 cable and 50 £2 termination.
b. Set DELAY TIME POS control to 1-00. Center
display.
c. ADJUST-Delay Start, R2782 (see Fig. 6-5), so that
second time mark is intensified.
d. Set DELAY TIME POS control to 9.00.
e. ADJUST-Delay Stop, R2748 (see Fig. 6-5), so that
10th time mark is intensified.
Fig. 6 5. Sweep adjustment locations.
6 10 REV. A, OCT 1975

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 455 and is the answer not in the manual?

Tektronix 455 Specifications

General IconGeneral
BrandTektronix
Model455
CategoryTest Equipment
LanguageEnglish

Related product manuals