EasyManua.ls Logo

Tektronix 455 - A and B Sweep Rate Accuracy

Tektronix 455
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Check455/A2/B 2
j. CHECKDisplay is triggered along negative-going
slope of waveform as B LEVEL control is rotated and is
not triggered (not visible) at either extreme of B LEVEL
rotation.
NOTE
I f the 11th time marker is n ot visible, set the A
TIM E/DIV switch one position counterclock
wise from the B TIM E/DIV switch. Example:
k. Disconnect test equipment.
A TIM E/DIV 1ms
B TIM E/DIV 0.5ms
21. A AND B SWEEP RATE ACCURACY
a. Set: A SOURCE NORM
B SOURCE STARTS AFTER
DELAY
A and B SLOPE OUT: +
HORIZ DISPLAY A
b. Connect test equipment as shown in Fig. 3-6.
c. CHECKA TIME/DIV accuracy according to Table
3-3. Incremental accuracy should be 1 time mark/division
within 2% (within 0.2 div at the 11th graticule line).
d. Set HORIZ DISPLAY to B DLY'D.
e. CHECKB TIM E/DIV accuracy according to Table
3-3 (adjust A LEVEL control as needed for a stable display).
TABLE 3-3
A and B Timing Accuracy
A and B
Time-Mark
TIME/DIV
Generator
Switch Setting
Output
.05 jus
50 ns
.1 ps
0.1 jus
.2 jus 0.2 jus
.5 jus
0.5 jus
1 jus
1 JUS
2 jus
2 jus
5 jus
5 jus
10 jus
10 jus
20 jus
20 jus
50 jus
50 jus
.1 ms 0.1 ms
.2 ms
0.2 ms
TG501 OR
EQUIVALENT
TIME-MARK
1907-13
Fig. 3-6. Time-mark generator test setup.
3-10
REV. A, OCT 1975

Table of Contents

Related product manuals