EasyManua.ls Logo

Tektronix 455 - Fig. 6-3. Crt and Calibrator Adjustment Locations

Tektronix 455
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjustments455/A2/B2
c. Set A TIME/DIV and VAR TIME/DIV controls to
obtain exactly one time mark/division (may be neces
sary to set A TIM E/DIV to .5 ms).
d. ADJUST-Y AXIS, R573 (see Fig. 6-3), to align
center time mark with center vertical graticule line.
e. INTERACTIONPosition display baseline to center
horizontal graticule line and readjust TRACE ROTA
TION control. Then recheck Y Axis alignment.
A6. GEOMETRY
a. Adjust VAR timing and horizontal POSITION con
trols to align time markers with vertical graticule lines.
Fig. 6-3. Crt and calibrator adjustment locations.
b. ADJUSTGeometry, R572 (see Fig. 6-3), for mini
mum bowing of time marks.
c. INTERACTIONBetween Geometry and Y Axis
Adjustments. Repeat both adjustments for best geom
etry and Y Axis alignment.
d. Disconnect time-mark generator.
A7. CALIBRATOR
a. Connect digital voltmeter leads to CALIBRATOR
output terminals.
b. Connect shorting jumper between TP376 and TP386
(see Fig. 6-3). If output voltage does not read between
+297 mV and +303 mV, continue with step c.
c. ADJUSTCalibrator Amplitude, R386 (see Fig. 6-3)
for +300 mV.
d. Remove shorting jumper between TP376 and TP386.
e. Connect test oscilloscope to CALIBRATOR output
jacks and check for 300 mV square wave, approximately
1 kHz.
f. Disconnect test oscilloscope.
REV. A, OCT 1975
6-5

Table of Contents

Related product manuals