455/A2/B2
VOLTAGE CONDITIONS
Voltages shown on this schematic diagram were measured with a Tektronix DM 501 Digital Multimeter. Voltage measurements can vary as
much as +20%.
No signal was applied to the vertical input. The Main Module INTENSITY control was set to midrange. The B2 HORIZ DISPLAY was set
to A INTEN and the B SOURCE switch to NORM.
WAVEFORM CONDITIONS
Waveforms below were taken from a Tektronix 7000-series oscilloscope equipped with readout, 7B series time-base, 7A13 Differential Com
parator and a 10X probe. The offset voltages were read directly from the 7A13 in the dc coupled mode. If a different test oscilloscope
system is used, the ac coupled mode may have to be used due to high dc offset voltages. This can cause some waveforms to appear slightly
different. Waveforms may vary as much as ±20%.
The Main Module calibrator signal was applied to the CH I vertical input through a compensated 10X probe. The Vertical Module CH 1
VOLTS/D IV switch was set to 5 mV. The test oscilloscope was triggered externally on the + slope from the (A) + GATE OUT connector
at the rear of the Main Module.
The B2 controls were set as follows:
A SOURCE NORM
B SOURCE CH 1
A and B COUPLING DC