EasyManuals Logo
Home>Tektronix>Pulse Generator>SPG600

Tektronix SPG600 User Manual

Tektronix SPG600
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
Getting Started
1-8 SPG600 & SPG300 Sync Pulse Generators User Manual
56. Press the left (
W) or right (X) arrow button to select Option and then press the
ENTER button.
57. Press the up (
S) or down (T) arrow button to select COLOR BAR.
58. Press the left (
W) or right (X) arrow button to select 100% Color Bars.
59. Connect the CH 5 connector to the waveform monitor using the 75
BNC
cable.
60. Set the waveform monitor to view the black burst signal.
61. Check that the waveform monitor displays the appropriate black burst signal.
62. Change the BNC cable from the CH 5 connector to the CH 6 connector.
63. Check that the waveform monitor displays the appropriate black burst signal.
64. Change the BNC cable from the CH 6 connector to the CH 7 connector.
65. Check that the waveform monitor displays the appropriate black burst signal.
66. Change the BNC cable from the CH 7 connector to the CH 8 connector.
67. Set the waveform monitor to view the color bars signal.
68. Check that the waveform monitor displays the appropriate color bars signal.
Signal Outputs (SPG600 Option 03 Only).
69. Perform steps 15 through 17.
70. Press the up (
S) or down (T) arrow button to select SDI OUTPUT (OPT).
71. Press the left (
W) or right (X) arrow button to select BLACK: 3 TEST: 4 and
then press the ENTER button.
72. Press the VIDEO button to enter the Video menu.
73. Press the up (
S) or down (T) arrow button to select TEST SIGNAL (SDI).
74. Press the left (
W) or right (X) arrow button to select Option and then press the
ENTER button.
75. Press the up (
S) or down (T) arrow button to select COLOR BAR.
76. Press the left (
W) or right (X) arrow button to select 100% Color Bars.
77. Connect the SDI 3 connector to the waveform monitor using the 75
BNC
cable.
78. Check that the waveform monitor displays the appropriate black burst signal.
79. Change the BNC cable from the SDI 3 connector to the SDI 4 connector.

Table of Contents

Other manuals for Tektronix SPG600

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix SPG600 and is the answer not in the manual?

Tektronix SPG600 Specifications

General IconGeneral
BrandTektronix
ModelSPG600
CategoryPulse Generator
LanguageEnglish

Related product manuals