TCP305 and TCP312 Performance Verification
5-30
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCP312 Current Probe Test Record
Photocopy this form and use it to record the performance test results.
TCP312 test record
Instrument Serial Number: Certificate Number:
Temperat ure: Relative Humidity %:
Date of Cal ibration: Technician:
Performance test Range, A/V
Test
current, A
Minimum Measured/calculated Maximum
DC Gain Accuracy (% Error) 1A/V 0.20 A -- 3 % +3%
0.60 A -- 3 % +3%
1.00 A -- 3 % +3%
10 A/V 1.20 A -- 3 % +3%
3.60 A -- 3 % +3%
6.00 A -- 3 % +3%
Rise Time 1A/V -- -- -- ≤3.5 ns
Bandwidth 1A/V .707 -- -- -- -- -- --