Specifications
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
4-5
Environmental Characteristics
The environmental characteristics in Table 4--6 are warranted performance
specifications. Unlike the warranted characteristics in Table 4--1, the environ-
mental characteristics are type tested; therefore there are no performance
verification procedures provided to test these characteristics. Unless otherwise
noted, these characteristics apply to all probes and amplifiers.
Table 4- 6: Environmental characteristics
Parameter Characteristic
Ambient Temperature
Operating 0 _Cto+50_C(32_F to 122 _F)
Nonoperating -- 4 0 _Cto+75_C(--40_F to 167 _F)
Humidity
Operating 5 to 95% R.H. to +30 _C(86_F)
5 to 85% R.H. +30 _Cto+50_C(86_F to 122 _F)
Nonoperating 5 to 95% R.H. to +30 _C(86_F)
5 to 85% R.H. +30 _Cto+75_C(86_F to 167 _F)
Altitude
Operating 2000 m (6800 ft) maximum
Nonoperating 12,192 m (40,000 ft) maximum
Random Vibration, Amplifiers
Operating 0.31 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Nonoperating 2.46 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Random Vibration, Probes
Operating 2.66 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Nonoperating 3.48 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Shock, Amplifiers 50 g, 11 m s duration, half-sine pulses
Electro-Magnetic Com pliance Meets FCC Part 15, Subpart B, Class A