TCP303 and TCP404XL Performance Verification
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5-43
TCP404XL Current Probe Test Record
Photocopy this form and use it to record the performance test results.
TCP404XL test record
Instrument Serial Number: Certificate Number:
Temperat ure: Relative Humidity %:
Date of Cal ibration: Technician:
Performance test Range
Test
current
Minimum Measured/calculated Maximum
DC Gain Accuracy (% Error) 1A/mV 3.0 A -- 3 % +3%
9.0 A -- 3 % +3%
15.0 A -- 3 % +3%
Rise Time 1A/mV -- -- -- -- -- -- 175 ns
Bandwidth 1A/mV -- -- -- 2MHz -- -- -- -- -- --