EasyManua.ls Logo

Tektronix TCPA300 SERIES - Removing the Outer Case and Internal Covers

Tektronix TCPA300 SERIES
205 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Disassembly
7-4
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Removing the Outer Case and Internal Covers
To replace internal components, you need to remove the outer case and internal
covers. You may also want to remove the case and covers to access the internal
adjustments. The following procedures describe the removal process:
1. Disconnect the AC cord and any probes or cables from the amplifier.
2. Remove the 3 screws from the rear of the amplifier and slide the case off the
rear side of the amplifier. See F igure 7--1.
Case
Amplifier
Screws (3)
Torque screws to 8 in-lbs
when reassembling
Figure 7- 1: Removing the case from the amplifier