TCP303 and TCP404XL Adjustments
6-20
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
6. If the DC gain adjustment was adjusted, verify the pulse response by
repeating the TCP303 and TCP404XL Transient Response and Coarse
Adjustment procedure beginning on page 6--13.
7. When the probe adjustments are done, reinstall the probe access plugs for the
1 s, 10 s, and DC Gain adjustments.