Reference
Run/Stop Butto
n. Push the Run/Stop button when you want the oscilloscope to
continuously acquire waveforms. Push the button again to stop the acquisition.
Single Button. Push the Single (single sequence) button when you want the
oscilloscope to acquire a single waveform and then stop. Each time you push the
Single button, the oscilloscope begins to acquire another waveform. After the
oscilloscope detects a trigger it completes the acquisition and stops.
Acquisition mode Single button
Sample, Peak Detect Sequence is complete when one acquisition is acquired
Average
Sequence is complete when the dened number of
acquisitions is reached; (See pag
e 75, Acquire.)
Scan Mode Display. You can use t he Horizontal Scan acquisition mode (also
called Roll mode) to continuously monitor signals that change slowly. The
oscilloscope displays waveform updates from the left to the right of the screen and
erasesoldpointsasitdisplaysnewpoints. A moving, one-division-wide blank
section of the screen separates the new waveform points from the old.
The oscilloscope changes to the Scan acquisition mode when you turn the
Horizontal Scale knob to 100 ms/div or slower, and select the Auto Mode option
in the Trigger Menu.
To disable Scan mode, push the Trig Menu button and set the Mode option to
Normal.
Stopping the Acquisition. While the acquisition is running, the waveform display
is live. Stopping the a cquisition (when you push the Run/Stop button) freezes
the display. In either mode, the waveform display can be scaled or positioned
with the vertical and horizontal controls.
Autorange
WhenyoupushtheAutoRange button, the oscilloscope activates or deactivates
the Autorange function. An LED light turns on adjacent to the AutoRange button
that indicates when the function is active.
This function automatically adjusts setup values to track a signal. If the signal
changes, the setup continues to track the signal. When you power on the
oscilloscope, autoranging is always inactive.
Options Comment
Autoranging
Activates or deactivates the Autorange function; when activated
the adjacent LED light turns on
Vertical and Horizontal Tracks and adjusts both axes
Vertical Only
Tracks and adjusts the Vertical scale; does not change the
horizontal settings
TDS2000C and TDS1000C-EDU Series Oscilloscope User Manual 77

Table of Contents

Other manuals for Tektronix TDS2000C series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2000C series and is the answer not in the manual?

Tektronix TDS2000C series Specifications

General IconGeneral
BrandTektronix
ModelTDS2000C series
CategoryTest Equipment
LanguageEnglish

Related product manuals