Appendix B: TPP0101 and TPP0201 Series 10X Passive Probes Information
1. Connect the pro
be to the oscilloscope channel that you plan to use for your
measurements.
2. Connect the pr
obe to the probe compensation output terminals on the
oscilloscope front panel.
WARNING. To avoid electric shock, only connect to the Probe Comp signal on the
oscilloscope when making this adjustment.
3. Push Autoset or otherwise adjust your oscilloscope to display a stable
waveform.
4. Adjust the trimmer in the probe until you see a perfectly at-top square wave
on the display. (See illustration.)
WARNING. To avoid electric shock, only use the insulated adjustment tool when
making compensation adjustments.
Connecting the Probe to the Circuit
Use the standard accessories included with the probe to connect to your circuit.
WARNING. To avoid electric shock when using the probe or accessories, keep
ngers behind the nger guard of the probe body and accessories.
To r
educe risk of shock, ensure the ground lead and ground spring are fully mated
before connecting the probe to the circuit under test.
116 TDS2000C and TDS1000C-EDU Series Oscilloscope User Manual

Table of Contents

Other manuals for Tektronix TDS2000C series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2000C series and is the answer not in the manual?

Tektronix TDS2000C series Specifications

General IconGeneral
BrandTektronix
ModelTDS2000C series
CategoryTest Equipment
LanguageEnglish

Related product manuals