Understanding Oscilloscope Functions
4. Continues to ac
quire data until the waveform record is full.
5. Displays the newly-acquired waveform.
NOTE. ForEdgeandPulsetriggers,theoscilloscope counts the rate at which
trigger events occur to determine trigger frequency. The oscilloscope displays
the frequency in the lower right corner of the screen.
Source
You can use the Trigger Source options to select the signal that the oscilloscope
uses as a trigger. The source can be the AC power line (available only with Edge
triggers), or any signal connected to a channel BNC or to the Ext Trig BNC.
Types
The oscilloscope provides three types of triggers: Edge, Video, and Pulse Width.
Modes
You can select the Auto or the Normal trigger mode to dene how the oscilloscope
acquires data when it does not detect a trigger condition. (See page 96, Mode
Options.)
To perform a single sequence acquisition, push the Single button.
Coupling
You can use the Trigger Coupling option to determine which part of the signal
will pass to the trigger circuit. This can help you attain a stable display of the
waveform.
To use trigger coupling, push the Trig Menu button, select an Edge or Pulse
trigger, and select a Coupling option.
NOTE. Trigger coupling affects only the signal passed to the trigger system. It
does not affect the bandwidth or coupling of the signal displayed on the screen.
To view the conditioned signal being passed to the trigger circuit, push and hold
down the Trig View button.
Position
The h orizontal position control establishes the time between the trigger and the
screen center. Refer to Horizontal Scale and Position; Pretrigger Information
for information on how to use this control to position the trigger. (See page 23,
Horizontal Scale and Position; Pretrigger Information.)
Slope and Level
The Slope and Level controls help to dene the trigger. The Slope option (Edge
trigger type only) determines whether the oscilloscope nds the trigger point on
the rising or the falling edge of a signal. The Trigger Level knob controls where
on the edge the trigger point occurs.
TDS2000C and TDS1000C-EDU Series Oscilloscope User Manual 21

Table of Contents

Other manuals for Tektronix TDS2000C series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2000C series and is the answer not in the manual?

Tektronix TDS2000C series Specifications

General IconGeneral
BrandTektronix
ModelTDS2000C series
CategoryTest Equipment
LanguageEnglish

Related product manuals