Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2- 261
Returns the number of points that are in the transmitted waveform
record. The num ber of points depends on DATa:STARt, DATa: STOP,
and whether DATa:SOUrce is YT or FFT. NR_Pt is at most 2500 for
YT and 1024 for FFT. NR_Pt is always at least one.
When the DATa:SOUrce is not displayed, the oscilloscope will
generate an error and return event code 2244.
Waveform
WFMPre:<wfm>:NR_Pt?
WFMPre:CH1:NR_Pt?
Returns <NR1>, which is the number of data points. If DATa:WIDth is
2, then there are <NR1>*2 bytes in the curve.
DATa:SOUrce, DATa:STARt, DATa:STOP
WFMPre:<wfm>:PT_Fmt
Same as WFMPre:PT_Fmt, except that <wfm> specifies the
source/destination waveform instead of DATa:SOUrce and
DATa:DESTINATION. For set commands, if <wfm> is not a
reference waveform, the oscilloscope generates error 2241.
WFMPre:<wfm>:PT_Off
Same as WFMPre:PT_Off, except that <wfm> specifies the
source/destination waveform instead of DATa:SOUrce and
DATa:DESTINATION.
For set commands, if <wfm> is not a reference waveform, the
oscilloscope generates error 2241.
Group
Syntax
Examples
Related Commands

Table of Contents