Page 4-10
©2004 TESA Switzerland All Rights Reserved.
User's Manual
MH3D
Movement is not smooth / Bearing is rubbing
• Stop using the machine immediately! Continued use could damage the guideway
system.
• Check to ensure that the air pressure for the machine is set to 55 psi.
• If the machine air pressure is set to 55 psi, consult TESA for help.
Oil appears from under the air bearings
• Clean the rails as described in “Cleaning the Bearing Surfaces”
• Check the machine’s air filters and replace the cartridge if necessary.
• If oil appears frequently, the air supply may be particularly poor. It may be
advisable to add a larger coalescing filter to the air supply upstream from the
machine.