3.7 Settings for the humidity sensor
Devices that comply with the "ASTM C191" standard are equipped with a sensor for measuring rel-
ative humidity.
Each sensor has its own parameter settings, which we configure on delivery. Our service or the
user must set and save the new parameters only if it should be necessary to change the sensor.
Figure 15: parameters of the humidity sensor
Parameters of the humidity sensor:
Slope
Zero voltage
Offset
The offset adjusts the value to a more accurate reference device. Normally, this value is set to „0.0 %rH“.
Press the button to get to the page for entering the slope.
Press the button to get to the page for entering the zero level.
Press the button to enable the use of the humidity sensor.
Ther Button changes to and when pressed again, you turn the sensor off.
You can observe the relative humidity on the page "monitoring" (Figure 30).
Slope
Zero level
Enable
Disable
settings> humidity sensor
Slope
Enable
Zero voltage
Offset