Single Station Vicat Test Apparatus
1.0360
24
3.7.3 Adjusting the offset of the humidity sensor
The offset of the humidity sensor adjusts the value of relative humidity of the sensor to a more accurate
referencing instrument.
rHRef = rHSens offset
Figure 18: Adjusting the offset of the humidity sensor
One-time short press of the button increases the value of the offset by 0.1 %rH.
Pressing and hold down this button increases the value continuously by 0.1 %rH.
If you press and hold the button for more than 3 seconds, the step size increases to 1 %rH.
This process continues until you either release the button or you reach the maximum permissible value.
One-time short press of the button decreases the value of the offset by 0.1 %rH.
Pressing and hold down this button decreases the value continuously by 0.1 %rH.
If you press and hold the button for more than 3 seconds, the step size increases to 1 %rH.
This process continues until you either release the button or you reach the minimum permissible value.
To get the set value also after switching off the device press the button.
Settings> Humidity sensor >Offset
%rH
-0.1