6 Using the product
39
6.2.3. Pipe volume
Based on the pipe volume calculation, unknown pipe volumes can
be determined and allocated to the relevant measuring location.
Particularly for the tightness test, where the stability time and
measurement time depends on the pipe volume, the advantage is
even more accurate and more reliable measurement.
Calling up the function:
> [] → Measurements → [OK] → Pipe volume → [OK].
Setting parameters
Parameter Selection Explanation
Gas feeding unit Yes/No Enable/disable gas feeding
unit
Test gas Air, CO2,
G30, G21,
propane,
natural gas,
Select test gas
1. Select parameter → [Edit].
2. Select or enter values: [▲], [▼] and sometimes [◄], [►] →
[OK].
Carrying out the measurement:
3. Start measurement: [].
> When Gas feeding unit Yes is selected: carry out steps in
accordance with the information on the display and confirm with
[OK].
- Zeroing starts (5 s)
- Volume calculation is carried out (bar shows progress of the
calculation)
-
Volume is displayed.
4. Adopt measurement result: [Apply].