6 Using the product
54
Setting parameters
Parameter Explanation
Stability time Set stability time
Measurement period Set measurement period
TRF 2012 stipulates a
measurement period of 10 min.
Test pressure (target) Set test pressure
According to TRF 2012, the test
pressure is 150 mbar.
1. Edit values: [Edit].
2. Enter values. [], [] and sometimes [], [] [OK].
Carrying out repeated testing PS>0.5 bar
1. Build up test pressure (target) manually with the test pump or a
compressor at the connection provided.
2. Start measurement: [].
- Zeroing starts (5 s)
Option
> End stability time early: [Next]
> End measurement period early: [Next]
- Measurement is cancelled.
- Pressure test is ended.
3. Evaluate measurement result: [], [] [OK].
- The measurement result is displayed
4 Save the measurement result: [Save].
Options
> [Options] Show Graphic: a section of past measuring
values is displayed in a line graph (zoom function available [],
[]).
> [Options] Manometer: The measuring values are displayed
in the same way.
> [Options] Factory setting: the values are reset to the
default settings.
> [Options] Overall graphic curve: the measuring values
across the overall measuring time are displayed in a line graph.

Table of Contents

Related product manuals