2 Safety
16 IFU, MIRUS Controller, en-GB, N.00
2.5.2 Positioning
WARNING Never use the MIRUS Controller when not in a horizontal position to
avoid the liquid agent entering the gas pathway.
If the MIRUS Controller has tipped over, wait for a minimum of
10 seconds to ensure no liquid is in gas pathways before applying
anaesthetic agent.
Ensure the MIRUS Controller is always in a stable position.
2.5.3 Risk of electrical hazard
CAUTION Use the enclosed original line supply cable only.
Make sure, that the device is not covered by any material (e.g. a
curtain) during the operation. This may interfere with the device’s
cooling system.
Operate the system only according to the given specification for
temperature and humidity. In case the system's temperature is higher
or lower than specified, allow the system to stabilise in the specified
operation temperature for one hour before operation.
Do not use the MIRUS System for intra or inner clinical transport.
Do not use the MIRUS System in an emergency / ambulance car.
Do not use the MIRUS System in a helicopter or airplane.
2.5.4 Risk of fire
CAUTION Do not use materials such as Ammonium, Phenol, or Acetone to clean
the device.
Do not use the MIRUS Controller when there is doubt in the proper
function of the electrical earth ground in the installation environment.
Do not use the MIRUS Controller in the presence of flammable
anaesthetics.