7 Alarms and messages
IFU, MIRUS Controller, en-GB, N.00 49
7.1.4 Patient alarms
Alarm
(in alphabetical
order)
Priority
Device
action
Set condition
Reset condition
Apnoea
Low
Alarm and
stop VA
delivery
No breathing activity.
Breathing activity
detected.
High et CO
2
High
Alarm
Measured et CO
2
>
max et CO
2
for three
consecutive breaths.
Measured et CO
2
within
the limit with the first
breath.
High et VA
VA = Sevo, Iso
or Des
High
Alarm
Measured et VA >
max etVA for three
consecutive breaths.
Measured et VA within
the limit with the first
breath.
Low et CO
2
High
Alarm
Measured et CO
2
<
min et CO
2
for three
consecutive breaths.
Measured et CO
2
within
the limit with the first
breath.
Low et VA
VA = Sevo, Iso
or Des
High
Alarm
Measured et VA < min
et VA for three
consecutive breaths.
Measured et VA within
the limit with the first
breath.
Low Vt
High
Alarm
Measured Vt
i
<
200 mL for five
consecutive breaths.
Measured Vt
i
≥ 200 mL
with the first breath.
High
Alarm and
stop VA
delivery
Measured Vt
i
< 50 mL
for five consecutive
breaths.
Measured Vt
i
≥ 50 mL
with the first breath.