3 Preparation
IFU, MIRUS Controller, en-GB, N.00 29
3.10 Changing alarm settings
Press menu button Alarm settings.
Note: Changing alarm settings is also
possible during application of VA.
Alarm settings
pre-set alarm limits for SEVO:
et CO
2
min = 20 mmHg
et CO
2
max = 50 mmHg
et SEVO min = 0.5 Vol%
et SEVO max = 2.5 Vol%
Apnoe time = 60 sec
Alarm vol. = 100 %
Refer to chapter 10.2.2 Alarm settings for
information about ISO and DES.
Press button for parameter to be changed,
e.g. et CO
2
max.
Button for selected parameter is light
grey.
Arrow keys are activated.
Use arrow keys to change parameter.