10 Technical specifications
IFU, MIRUS Controller, en-GB, N.00 61
10.2 Controls and Ranges
10.2.1 Agent dosage
For
Values
Isoflurane
Range: MAC 0.1 to 2.0 @ MVi 3.0 to
15.0 L/min where
MAC 1 = 1.15 Vol% @ age = 40
Increment: 0.1 MAC @ single taps on button
0.5 MAC @ pressing button > 1 sec
Sevoflurane
Range: MAC 0.1 to 2.0 @ MVi 3.0 to
15.0 L/min where
MAC 1 = 1.9 Vol% @ age = 40
Increment: 0.1 MAC @ single taps on button
0.5 MAC @ pressing button > 1 sec
Desflurane
Range: MAC 0.1 to 2.0 @ MVi 3.0 to
15.0 L/min where
MAC 1 = 6.7 Vol% @ age = 40
Increment: 0.1 MAC @ single taps on button
0.5 MAC @ pressing button > 1 sec
10.2.2 Alarm settings
For
Values
et CO
2
min. / et CO
2
max.
Range: 15 to 150 mmHg / 2.0 to 19.7% / 2.0 to
20.0 kPa
Increment: 1 mmHg / 0.5% / 0.5 kPa @ single
taps on button
5 mmHg / 1.0% / 1.0 kPa @ pressing
button > 1 sec
et ISO min. / et ISO max.
Range: 0.0 to 6.0 Vol%
Increment: 0.1 Vol% @ single taps on button
0.5 Vol% @ pressing button > 1 sec
et SEVO min. / et SEVO max.
Range: 0.0 to 8.0 Vol%.
Increment: 0.1 Vol% @ single taps on button
0.5 Vol% @ pressing button > 1 sec
et DES min. / et DES max.
Range: 0.0 to 18.0 Vol%
Increment: 0.1 Vol% @ single taps on button
0.5 Vol% @ pressing button > 1 sec
Apnoea time
Range: 15 to 60 sec
Increment: 5 seconds
Alarm volume
Range: 50 to 100%
Increment: 10% @ single taps on button