8 Standard values
IFU, MIRUS Controller, en-GB, N.00 55
8 Standard values
8.1 Alarm settings and application defaults
Screen
Parameter
Default value
Home
MAC
0.5
Patient data
Gender
male
Age
40 years
Ideal body weight
80 kg
Body height
180 cm
Vol.
SEVO 0.9 Vol%
ISO 0.6 Vol%
DES 3.4 Vol%
MAC Pilot
Wash-in speed
medium
Alarm settings
et CO
2
min
20 mmHg
et CO
2
max
50 mmHg
et VA min
ISO 0.3 Vol %
SEVO 0.5 Vol %
DES 1.7 Vol%
et VA max
MAC * 2.0
(VA specific)
Apnoea time
60 seconds
Alarm volume
100%