12 High priority alarms
66 IFU, MIRUS Controller, en-GB, N.00
12 High priority alarms
Alarm
What MIRUS Controller does
What to check or to do
Device
inclined!
Device pauses the application and
closes all valves until device is
positioned on a level surface.
1. Reposition device on a level
surface.
Dosage failed!
Start VA?
Device stops the application.
1. Restart VA delivery by
activating Play and Confirm or
turn off device.
Failure!
Turn off.
Device stops the application.
Turn off tag becomes selectable.
A major technical failure was
recognised.
1. Turn off device into OFF mode
(Standby).
2. Wait 5 seconds and try to start
device again.
3. If during self test "Sorry…"
message appears the device is
seriously damaged. Call your
service organisation.
Fill port flap
still open!
Continues working.
1. Close Fill port flap.
Heating
system failed.
Continues working.
1. Turn off device into OFF mode
(Standby).
High et VA
Continues working.
1. Patient okay (CO
2
, SpO
2
)?
2. Set MAC okay?
3. Alarm settings okay?
High et CO
2
Continues working.
1. Ventilator settings okay?
2. Alarm settings okay?
3. Patient okay?
Interface
disconnected
Device pauses the application.
1. Reconnect MIRUS Interface
plug.
Low et CO
2
Continues working.
1. MIRUS Interface okay?
2. Ventilator settings okay?
3. Alarm settings okay?
Low et VA
Continues working.
1. Delivery active (Start button!)?
2. Alarm settings correct?
3. MRUS Interface okay?
Low Vt
Continues working.
1. Patient okay?
2. Ventilator settings okay?
Vt has to be ≥ 200mL!
Occlusion of
Interface
Device pauses the application.
1. MIRUS Interface bent or
pinched?
2. Fresh gas flow of Anaesthesia
Work Station minimum
1.5 * MV?
Reservoir
empty
Continues working.
1. Refill reservoir promptly.
UPS battery
low
Continues working.
1. Allow recharging of battery.