12 High priority alarms
66 IFU, MIRUS Controller, en-GB, N.00
12 High priority alarms
What MIRUS Controller does
Device pauses the application and
closes all valves until device is
positioned on a level surface.
1. Reposition device on a level
surface.
Device stops the application.
1. Restart VA delivery by
activating Play and Confirm or
turn off device.
Device stops the application.
Turn off tag becomes selectable.
A major technical failure was
recognised.
1. Turn off device into OFF mode
(Standby).
2. Wait 5 seconds and try to start
device again.
3. If during self test "Sorry…"
message appears the device is
seriously damaged. Call your
service organisation.
Fill port flap
still open!
1. Turn off device into OFF mode
(Standby).
1. Patient okay (CO
2
, SpO
2
)?
2. Set MAC okay?
3. Alarm settings okay?
1. Ventilator settings okay?
2. Alarm settings okay?
3. Patient okay?
Device pauses the application.
1. Reconnect MIRUS Interface
plug.
1. MIRUS Interface okay?
2. Ventilator settings okay?
3. Alarm settings okay?
1. Delivery active (Start button!)?
2. Alarm settings correct?
3. MRUS Interface okay?
1. Patient okay?
2. Ventilator settings okay?
Vt has to be ≥ 200mL!
Device pauses the application.
1. MIRUS Interface bent or
pinched?
2. Fresh gas flow of Anaesthesia
Work Station minimum
1.5 * MV?
1. Refill reservoir promptly.
1. Allow recharging of battery.