3 Preparation
22 IFU, MIRUS Controller, en-GB, N.00
3.5 Filling during start-up sequence
If liquid volume level in reservoir is less than 75 mL, MIRUS Controller will require filling prior
to the system test. Wait for a request to fill (message on screen) and follow the instructions.
Note: For further information about the bottle adapter refer to the manufacturer's instructions.
CAUTION Only the suitable bottle adapter will match with MIRUS Controller.
CAUTION Don’t force an unsuitable adapter into the fill port.
CAUTION Double check bottle adapter colour, fill port collar colour and agent
bottle colour to match dedicated volatile anaesthetic agent prior to
filling.
WARNING Do not press the fill port valve manually down at any time. This will
result in pressure and vapour release.
WARNING Slowly remove bottle from fill port to reduce vapour pressure to avoid
ambient pollution.
WARNING Avoid spilling volatile anaesthetic agent during filling process to avoid
ambient pollution.
Perform
CAUTION Do not try to open the
flap manually, it could
be seriously damaged.
Press Fill port flap button to open fill port
flap or
press Turn off button to shut down device.