10 Technical specifications
IFU, MIRUS Controller, en-GB, N.00 63
10.3.3 Gas monitoring accuracy
Specification
Values
CO
2
0 1 % ± 0.1%
1 5 % ± 0.2%
5 7 % ± 0.3%
7 10 % ± 0.5%
> 10 % unspecified
Isoflurane
0 1 % ± 0.15%
1 5 % ± 0.2%
> 5 % unspecified
Sevoflurane
0 1 % ± 0.15%
1 5 % ± 0.2%
5 8 % ± 0.4%
> 8 % unspecified
Desflurane
0 1 % ± 0.15%
1 5 % ± 0.2%
5 10 % ± 0.4%
10 15 % ± 0.6%
15 18 % ± 1.0%
> 18 % unspecified
10.4 Monitoring system
Specification
Device
Safety monitoring
Patient
Breathing activity
Alarms
Apnoea
Tidal volume
Low Vt
Agent dosage monitoring
Dosage start
start of inspiration
Dosage end
end of inspiration or end of delivered anaesthetic
agent volume
Control
Monitoring of applied volume of anaesthetic
agent. Comparison of calculated with measured
volume
Alarms
Dosage failed! No delivery
Gas monitoring
Agent
Isoflurane, Sevoflurane, Desflurane
Numerical data
et VA
Alarms
Low et VA, High et VA
Metabolic gases
CO
2
Numerical data
et CO
2
Alarms
Low et CO
2
, High et CO
2