L
L
e
e
a
a
k
k
T
T
e
e
s
s
t
t
i
i
n
n
g
g
To ensure the integrity of the radioactive source encapsulation,
the 3241 gauges must be leak tested at intervals not exceeding
six months unless otherwise indicated by your license. Some
countries may require leak testing be performed only by
authorized, licensed institutions. Troxler is approved to perform
leak test services. Call 1-877-TROXLER to order a leak test kit.
Each individual who leak tests the gauge is required to wear
appropriate personnel monitoring during this duty.
Using the Troxler leak test kit and accompanying instructions
perform the following procedure:
1. Write the date and gauge model and serial numbers on the
form and on the wipe folder.
2. Open the sample chamber door and locate the round
indentation on the inside top plate (centered between the
two screw heads).
3. Using the wipe disk, wipe the area around the indentation.
4. Pack the disk, as instructed, in the envelope and mail to
Troxler for analysis.
5. Secure the gauge properly.