Use the Recall function to view the results of the last reading
taken with the gauge. Press
〈SHIFT〉 and 〈1〉 for the Recall
display:
T
T
a
a
r
r
g
g
e
e
t
t
P
P
r
r
e
e
c
c
i
i
s
s
i
i
o
o
n
n
If you require a higher precision, the gauge may be programmed
to automatically increase the count time to reach the desired
precision. Using an ASTM precision equation, the count time will
be adjusted. The maximum time allowed under this function is
60 minutes and is rounded to the nearest 15-second interval.
Press 〈SHIFT〉 and 〈2〉 for the Precision display:
Input the required precision and press 〈ENTER〉. The gauge will
request the %AC. Input the %AC and press
〈ENTER〉. The gauge
calculates the time required to obtain the requested precision.
The gauge automatically enables the count time calculated to
reach the target precision.
Tar get Pr eci si on
_____