Model 3241 Series
7-11
To execute a Stat Test, place the gauge in an area free of other
gauges or radioactive sources. Press
SHIFT and 9 to display
the Special menu.
Press 1 for the display:
To accept the displayed count time, press NO.
To change the count time, press YES and enter the desired
count time. Select a count time of 1, 4, 8 or 16 minutes and then
press
ENTER.
Remove any samples or sample pans from the gauge. Close the
door and press
START.
After the 20 counts have been completed, the gauge will display
the average counts and indicate if the test passed or failed. The
individual count data may be viewed or printed at this time by
pressing
YES at the view data prompt. To return to the Ready
mode without viewing the data, press
NO. View the data by
pressing
YES at the view data prompt.
Select the method of viewing the data. If a printout is desired
refer to page 7-14 for the location of the serial port and baud
rate selection. A sample printout is shown on page 7-5.
- STAT TEST-
Ti me:
xx
mi n.
Do
you
want
t o
change?

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler 3241 Series and is the answer not in the manual?

Troxler 3241 Series Specifications

General IconGeneral
BrandTroxler
Model3241 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals