To execute a Stat Test, place the gauge in an area free of other
gauges or radioactive sources. Press
〈SHIFT〉 and 〈9〉 to display
the Special menu.
Press 〈1〉 for the display:
To accept the displayed count time, press 〈NO〉.
To change the count time, press 〈YES〉 and enter the desired
count time. Select a count time of 1, 4, 8 or 16 minutes and then
press
〈ENTER〉.
Remove any samples or sample pans from the gauge. Close the
door and press
〈START〉.
After the 20 counts have been completed, the gauge will display
the average counts and indicate if the test passed or failed. The
individual count data may be viewed or printed at this time by
pressing
〈YES〉 at the view data prompt. To return to the Ready
mode without viewing the data, press
〈NO〉. View the data by
pressing
〈YES〉 at the view data prompt.
Select the method of viewing the data. If a printout is desired
refer to page 7-14 for the location of the serial port and baud
rate selection. A sample printout is shown on page 7-5.