4-10
Place the gauge on the test material. To begin taking the four one-
minute counts, press
START/ENTER. After each count is
complete, the operator must initiate the next count by pressing
START/ENTER.
To create a partial calibration and return to the Ready mode, press
OFF/NO. To complete the special calibration by entering the
density, press
ON/YES.
To change the value of a digit, press the up and down arrows. To
select the next digit, press
START/ENTER. After the density
value is entered, the special calibration routine readjusts the gauge
for the new material and indicates that the special calibration is
enabled. The special calibration is only valid for the depth selected
during the special calibration.
Press START for
Reading # x
Enter known
density now?
Density: xx
( or ENTER)
SPECIAL CALIB.
ENABLED!