EasyManuals Logo

Troxler 3430 User Manual

Troxler 3430
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
Model 3430
C-5
Check the output voltage of your charger. The correct output
voltage is indicated on the charger unit.
Check that you are using the correct charger.
Remove any loose screws or foreign objects from the gauge
interior that may cause an electrical short to ground.
The AC charger may be defective. Check voltage output of
charge with a voltmeter, or use the DC charger to charge the
batteries.
SATISFACTORY COUNTS, BUT RESULTS ARE IN ERROR
Ensure the measurement depth corresponds to the actual source
rod depth.
Check calibration constants.
Check to see if an offset (density, moisture, trench or special) is
enabled.
Ensure that the standard counts are correct.
Ensure that the index rod is seated in bottom of notch.
MAINTENANCE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler 3430 and is the answer not in the manual?

Troxler 3430 Specifications

General IconGeneral
Measurement PrincipleNuclear
Power SourceRechargeable battery
DisplayLCD
Data StorageInternal memory
TypePortable
Material MeasuredSoil, Asphalt
Moisture Measurement Range0 to 60%
Operating Temperature-10 to 50°C
SourceCesium-137

Related product manuals