Follow the instructions provided with the internal angle device
to measure the bottom angle. Ensure that the probes of the
internal angle device are aligned along Position 1 with the edge
of the mold pull as shown in Figure 41.
6
6
.
.
Remove the mold from the chamber and record the measured
angle from the internal angle device (if you use a device with
recording capabilities).
7
7
.
.
Repeat the two previous steps to obtain a second bottom angle
measurement where the device is aligned along Position 2.
8
8
.
.
Follow the instructions provided with the internal angle device
to measure the top angle. Ensure that the probes of the internal
angle device are aligned along Position 1 with the edge of the
mold pull as shown in Figure 41.
9
9
.
.
Remove the mold from the chamber and record the measured
angle from the internal angle device.
1
1
0
0
.
.
Repeat the two previous steps to obtain a second top angle
measurement where the device is aligned along Position 2.
1
1
1
1
.
.
Calculate the average of the bottom angle and top angle
measurements (four measurements total).
1
1
2
2
.
.
Compare the average to the entered angle of 1.16. If the
difference is more than ±0.03°, refer to the Angle Offset section
on page 412.