Model 5850v2 49
4. CALIBRATION
CAUTION
Use caution when opening and closing the chamber door
to prevent catching or pinching the load cell cables.
5
5
.
.
Press ENTER. The compactor measures the unloaded
pressure on the load cell. During this measurement, the control
unit displays:
Pressure Cal/Verify
Testing Calibration
Load Cell
Please Wait
NOTE
If the load cell is not properly connected, the control
unit briefly displays the following error message. Check
the load cell connection. Press ENTER to continue.
Pressure Cal/Verify
Communication Error
Check Load Cell
Press <ENTER>
6
6
.
.
After measuring the unloaded pressure on the load cell, the
compactor raises the ram until the load cell contacts the upper
puck plate. The compactor then applies pressure and displays:
Pressure Cal/Verify
Applying Pressure
During calibration and verification, the compactor raises the
compaction ram and performs a series of pressure measurements at
high and low pressure. Upon completion, the compactor verifies the
pressure calibration, and then displays the compaction pressure as
measured internally by the compactor (System) and as measured by
the load cell (Load Cell):