4.2 Water quantity for a cycle
The amount of water in the autoclave chamber necessary for each
sterilization cycles as follows:
1730 2340/2540 3545 3850 3870
350
ml.
12
ozs.
450
ml.
16
ozs.
700
ml.
25
ozs.
850
ml.
30
ozs.
1000
ml.
35
ozs.
It is imperative to have the correct amount of water for proper
operation of the autoclave!
4.3 Lifting and Carrying
Caution:
Before moving the autoclave, make sure that the electric cord is
disconnected from the power source and that there is no pressure in
the chamber.
1. Disconnect the power supply cord.
2. Drain the water from the reservoir and vessel.
Lifting and carrying should be done by two people.
Do not drop the device!
4.4 Loading and Unloading the Device
Protective equipment and clothes should be implemented in accordance
to local and national regulations and/or rules!