11.6 Temperature Safety Thermostat
(Located on the rear side of the heaters)
The autoclave is supplied with a temperature thermostat that maintains the
temperature during the dry stage, by connecting and disconnecting the
electric power.
This device automatically disconnects the heating elements in case of a rise
in temperature.
The electric power is automatically reconnected when the chamber cools
down.
♦ To replace this safety thermostat, remove the rear cover, unscrew the
thermostat and replace it.
11.7 Raising the Working the Temperature of the Safety Thermostat
Caution:
Only authorized technician should do this operation!
1. Unscrew the rear cover of the autoclave.
2. With a screwdriver, turn the central screw slightly clockwise to raise the
temperature.
3. Replace the rear cover.
Safety Thermostat