11.6 Temperature Safety Thermostat
(Located on the rear side of the heaters)
The autoclave is supplied with a temperature thermostat that maintains the
temperature during the dry stage, by connecting and disconnecting the
electric power.
This device automatically disconnects the heating elements in case of a rise
in temperature.
The electric power is automatically reconnected when the chamber cools
down.
To replace this safety thermostat, remove the rear cover, unscrew the
thermostat and replace it.
11.7 Raising the Working the Temperature of the Safety Thermostat
Caution:
Only authorized technician should do this operation!
1. Unscrew the rear cover of the autoclave.
2. With a screwdriver, turn the central screw slightly clockwise to raise the
temperature.
3. Replace the rear cover.
Safety Thermostat