7. OPERATION
To avoid possible damage, do not leave the autoclave un-attended while in
operation
1. Ensure that the drain valve is in a CLOSED position.
2. Remove the water reservoir cover.
3. Pour distilled or mineral-free water into the reservoir through the opening
on top of the autoclave until it reaches the base of the safety valve holder
(the base must no immerse in the water.
3.1 Verify that the water quality is according to the requirement specified
in para. 7.
Caution
Under no circumstances should water be filled above the Safety Valve
Holder.
Use distilled or mineral-free water only. Tap water may block the hole of the
air trap jet. This causes air pockets that prevent the temperature from rising. It
is essential that from time to time during the sterilization period, a spray of
steam should escape, causing a hissing sound. If no steam is evident, follow
instructions for air trap jet cleaning procedure (para. 11.2).
Caution:
Daily before operation, check the water level in the reservoir and add water
when required. Once a week or after 20 cycles (the shorter period) replace
the water in the reservoir.
4. Insert the plug into the socket at the rear of the autoclave.
5. Insert the plug into the electrical source.
6. Turn the main switch to START position.