EasyManua.ls Logo

Tuttnauer 3545 - 5 Water Quality

Tuttnauer 3545
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 14 of 63 Pages
5. WATER QUALITY
The distilled or mineral – free water supplied to the autoclave should have the
physical characteristics and maximum acceptable level of contaminants
indicated in the table below:
Physical characteristics and acceptable contaminants levels in water,
for sterlizers
Evaporate residue
15 mg/l
Silica
2 mg/l
Iron
0.2mg/l
Cadmium
0.005 mg/l
Lead
0.05 mg/l
Rest of heavy metals
0.1 mg/l
Chloride
3 mg/l
Phosphate
0.5 mg/l
Conductivity
50 µs/cm
pH 6.5 to 8
Appearance Colourless, clean, without sediment
Hardness
0.1 mmol/l
Compliance with the above data should be tested in accordance with
acknowledged analytical methods, by an authorized laboratory.
Attention:
We recommend testing the water quality once a month. The use of water for
autoclaves that does not comply with the table above may have severe impact
on the working life of the sterilizer and can invalidate the manufacturer’s
guarantee.

Related product manuals