11.2 Dismantling the Outer Cover of the Autoclave.
Caution:
Allow the instrument to cool before removing the outer covers.
Warning:
Before starting disconnect the instrument from the power source and
make sure there is no pressure in the autoclave.
Then proceed as follows:
1. Remove the screws holding the rear cover (1).
2. Remove the screws holding the cover to the base (2).
3. Pull the cover upwards.