Page 16 of 63 Pages
11. Verify that the packaging method is in accordance with good practice
approach and the packaging materials are in accordance with the
applicable standards (e.g. EN868 series).
12. Place a sterilization indicator strip in each tray.
13. Place instruments with ratchets opened and unlocked or clipped on the
first ratchet position.
14. Disassemble or sufficiently loosen multiple-part instruments prior to
packaging to permit the sterilizing agent to come into contact with all
parts of the instrument.
15. Tilt on edge items prone to entrap air and moisture, e.g. hollowware, so
that only minimal resistance to air removal, the steam passage and
condensate will be met.
16. Load items within the boundaries of the tray so that they do not touch
the chamber walls, or fall off when the loaded car is inserted into the
autoclave.
17. The operator may use racks to allow for adequate separation of packaged
instruments.
18. Load trays loosely to capacity.
19. Once a week, use a biological spore test indicator in any load to make
sure sterilization is performed.
20. Make sure that all instruments remain apart during the sterilization cycle.
21. Empty canisters should be placed upside-down, in order to prevent
accumulation of water.
22. Allow a distance of approximately 2.5 cm (1”) between trays to permit
steam circulation.
23. Wrapped Instruments
Wrapped instruments should be packed in material that promotes drying
such as autoclave bag, autoclave paper, and muslin towels.
It is highly recommended to utilize the Tuttnauer Pouch Rack. This
rack allows the operator to place pouches on their side, thus increasing
the capacity of the autoclave significantly and promoting better drying of
the instruments. Contact your dealer for details.
Note:
A table “Suitability of steam sterilization processes for various goods and
method of packing” is added to the accompanying documents.
Pouch Rack