9.11 Working with power ON in Service Area (Autoclaves with sliding doors)
Warning!
It is strictly forbidden to enter the service area!
If maintenance is necessary behind any service access panel(s) while the system is
operating, proceed as follows:
1. Ensure the chamber door(s) are closed.
2. Lock the system by pressing the emergency switch to turn OFF all electrical
power to the autoclave and withdraw the key.
3. Open the service access panel(s) and enter the service area, and take the key
with to prevent accidents and unauthorized use.
4. Disconnect the door opening solenoid(s).
5. Re-insert the key, and turn ON the all electrical power to the autoclave to
operate the system.
6. Carry out the required maintenance.
7. Turn OFF all electrical power to the autoclave once again. See step 2 & 3.
8. Re-connect the door opening solenoid(s).
9. Replace the service access panel(s).
10. Re-insert the key, and turn ON the all electrical power to the autoclave, and
continue working.