3.8 Water Quality
Physical Characteristics and Maximum acceptable contaminants levels in
water or steam, for steam generator and sterilizers
(According to EN 285:2006).
Contaminants in
water supplied to
generator
Contaminants in
condensate at
steam inlet to
sterilizer
Rest of heavy metals
except iron, cadmium,
lead
pH value (degree of
acidity)
Hardness (Σ ions of
alkaline earth)
Colorless, clean, without sediments
Compliance with the above data should be tested in accordance with acknowledged
analytical methods, by an authorized laboratory.
Attention:
The use of water for autoclaves that do not comply with the table above may have
severe impact on the working life of the sterilizer and can invalidate the
manufacturer’s guarantee.
Tap Water supply
Maximum hardness value 0.7-2.0 mmol/l
Note:
The use of soft water is strictly forbidden!
Please consult a water specialist!