MAN205-0195042EN Rev. A Aug. 2011
Page 75 of 137 pages
Note:
When a sterilization program is running, the LED RUN lights up on the control
panel.
5. At the end of the cycle a buzzer sounds and the word "END" is displayed.
Press STOP to cancel the buzzer. Open the door by pressing the F3 key while
"OPEN" is displayed above this key.
6. Take out the sterilized material from the chamber following the steps
mentioned in the ‗Un-loading‘ paragraph.
8.1 Failed Cycle
In case of failed cycle or manual stop by operator, the two LED's RUN and FAIL are lit.
The reason for failure is displayed on the screen. At the end of the cycle an interrupted
sound is heard. Press the STOP key to cancel the alarm.
1. Press STOP.
2. Solve the problem.
3. Log the displayed message in the autoclave log book or any other appropriate
document.
4. If you cannot solve the problem, turn OFF the autoclave and call for service.
Note:
Verify that the pressure in the chamber is atmospheric pressure. Do not attempt
to open the door if the pressure in the chamber is above atmospheric pressure.
If cycle failed, the front door only can be opened.