MAN205-0195042EN Rev. A Aug. 2011
Page 13 of 137 pages
Electrical Power Connection
Warning!
Only a qualified electrician may perform the electrical connections!
The electrical units should not be placed near water sources.
Connect the power cord to the electric box of the autoclave
Electrical data
3 Ph, 208V/ 60Hz, 105 A
Electrical cable requirements
5 X 50 mm
2
Note:
Before operating the vacuum pump, make sure that the pump shaft is free and
you can turn it with a screw driver.
Warning!
It is strictly forbidden to enter the autoclave chamber!
If, for any reason (for cleaning, maintenance or if something falls down), it is
necessary to enter the chamber, the person must lock the system by pressing the
emergency switch and take the key with him to prevent accidents and
unauthorized use.
Warning!
It is strictly forbidden for any person to enter the service area behind the service
access panels, except the trained technician.
If, for any reason it is necessary to open the service access panel, the person
must shut the system by pressing the emergency switch, withdrawing the key and
keeping the key on him, to prevent accidents and injuries.
The technician must follow the instructions in the SERVICE INSTRUCTIONS,
Working with power ON in Service Area” Section.